Syringe Filters Solvent Compatibility Chart
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
ACIDS | ||||||
Acetic, 5% | L | R | R | R | R | R |
Acetic, 10% | L | R | R | R | R | R |
Acetic, 25% | N | L | R | R | R | R |
Acetic, Glacial | N | N | R | R | R | R |
Boric | – | L | – | R | – | – |
Formic 25% | L | N | – | R | – | R |
Hydrochloric 15% | L | L | R | R | L | R |
Hydrochloric 25% | N | N | R | R | – | R |
Hydrochloric concentrated | N | N | L | R | N | R |
Hydrofluoric 10% | N | N | – | – | – | – |
Hydrofluoric 35% | N | N | – | R | – | – |
Nitric 25% | N | N | R | R | – | R |
Nitric 6N, 38% | N | N | L | R | R | R |
Nitric concentrated | N | N | N | R | N | R |
Phosphoric 25% | L | N | R | R | – | R |
Sulfuric 25% | N | N | N | R | – | R |
Sulfuric 6N, 29% | N | N | N | R | – | R |
Sulfuric concentrated | N | N | N | R | N | R |
Trichloroacetic 10% | N | N | – | R | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
ALKALINES | ||||||
Ammonium Hydroxide 25% | N | R | R | R | L | R |
Formalin 30% | L | L | R | – | – | – |
Sodium Hydroxide 3N, 12% | N | R | R | R | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
ALCOHOLS | ||||||
Amyl Alcohol | L | R | N | R | R | R |
Benzyl Alcohol | L | L | L | L | L | R |
Butyl Alcohol | L | R | L | R | R | R |
Butyl Cellosolve | N | L | – | L | – | – |
Ethanol 70% | L | R | L | R | R | R |
Ethanol 98% | N | R | N | R | R | R |
Ethylene glycol | L | R | R | R | R | R |
Glycerol | L | R | R | R | R | R |
Isobutyl Alcohol | L | L | L | L | L | R |
Isopropanol, n-Propanol | L | R | R | R | R | R |
Methanol 98% | N | R | L | R | R | R |
Methyl Cellosolve | L | L | – | L | – | – |
Propylene glycol | L | – | R | R | R | R |
Phenol, Aqueous 10% | – | R | – | R | R | – |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
HYDROCARBONS | ||||||
Hexane | L | R | L | R | R | N |
Xylene | L | R | N | R | R | – |
Kerosene, Gasoline | L | R | R | R | R | L |
Tetrakin, Decalin | N | R | – | R | R | – |
Toluene, benzene | L | R | N | R | R | N |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
HALOGENATED HYDROCARBONS | ||||||
Carbon Tetrachloride | N | N | N | N | N | L |
Chloroform | N | N | N | R | R | L |
Methylene Chloride | N | L | N | R | N | L |
Monochlorobenzene | N | N | – | R | – | R |
Trichloroethylene | N | N | N | R | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
KETONES | ||||||
Acetone | N | R | N | R | N | R |
Cyclohexanone | N | L | N | R | N | R |
Isopropylacetone | – | R | – | R | N | – |
Methyl Ethyl Ketone | N | R | N | R | N | R |
Methyl Isobutyl Ketone | N | R | – | R | N | L |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
ESTERS | ||||||
Amyl Acetate | N | R | L | R | – | L |
Amyl Propyl & Butyl Acetate | L | – | – | R | – | L |
Benzyl Benzoate | – | – | – | R | – | – |
Butyl Acetate | N | – | N | – | – | L |
Ethyl Acetate & Methyl Acetate | N | R | N | R | R/L | L |
Isopropyl Myristate | – | – | – | R | – | – |
Methyl Cellosolve Acetate | N | – | – | R | – | R |
Propylene Glycol Acetate | – | – | – | R | – | R |
Tricresyl Phosphate | – | – | – | R | – | – |
Isopropyl Acetate | L | R | – | – | R | – |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
OXIDES - ETHERS | ||||||
Dimethylsulfoxide (DMSO) | N | R | N | R | N | R |
Dioxane & Tetrahydrofuran | N | R | L | R | L | R |
Ethyl Ether | L | R | R | R | R | L |
Isopropyl Ether | – | – | – | R | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
SOLVENTS WITH NITROGEN | ||||||
Acetonitrile | N | R | N | R | N | R |
Aniline | N | – | – | R | – | – |
Diethylacetamide | N | L | N | R | N | – |
Dimethyl Formamide | N | R | N | R | N | R |
Pyridine | N | R | N | R | – | N |
Triethanolamine | – | R | – | R | N | – |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF | PP |
MISCELLANEOUS | ||||||
Formaldehyde Solution 30% | L | R | R | R | R | R |
Hydrogen Peroxide 30% | N | L | N | R | R | – |
Pyridine | N | R | N | R | R | – |
Silicone Oil & Mineral Oil | R | R | R | R | R | R |
R = | Recommended. No significant change observed in flow rate or bubble point of the membrane, nor visible indication of chemical attack. |
L = | Limited Recommended Use. Moderate changes in physical properties. The filter may be suitable for short term, noncritical use. |
N = | Not Recommended. The membrane may be unstable. |
– = | Insufficient Data. Information is not available. Trial testing is recommended. |
Top 7 Reasons to Use Restek Syringe Filters![syr_filters_ph_so_spe.jpg](https://content.restek.com/content/published/api/v1.1/assets/CONTC86804D0447E4388A00F9646E18EA496/native?download=false&channelToken=6428daeb0e4e4341beecd8e2b41b4d46)
- Protect any analytical system.
- Extend LC column lifetime.
- Achieve more reproducible analyses.
- Variety of membranes, porosities, and diameters available.
- Luer lock inlet provides strong, leak-tight syringe connection to withstand filtration pressure.
- Rugged polypropylene construction— autoclavable to 121 °C for 15 minutes (75 psi).
- Labeled (13, 25, and 30 mm, only) and color coded for easy identification.