Restek
Home / Resource Hub / ChromaBLOGraphy / Maximum temperatures of packed columns Hydrocarbon Analysis

Maximum temperatures of packed columns – Hydrocarbon Analysis

21 Jun 2015

For my third post in this series, I would like to focus on specialty packed columns for hydrocarbon analysis.  In some cases, the packings in these columns are proprietary, so detailed information cannot be provided.  In other cases, application information may be limited.  However, we can provide maximum temperature limits (see table below).

For other columns used for hydrocarbon analysis, you should be able to find the maximum packing temperature based upon the maximum temperature of the liquid phase(s) and/or the porous polymer(s).

If there is a packed column for hydrocarbon analysis (which is sold by Restek) and not included in this table or in either link above, contact Restek. Thank you.

Column and/or Packing Max Temp (°C)
D3606 Column Set 140
0.19% picric acid on CarboBlack C 120
23% Rt-1700 on Chromosorb PAW 110
n-Octane on Res-Sil C 150
OPN on Res-Sil C 150
2abc Refinery Gas Column Set 110/200*
Alumina F-1 300
Rtx-1 SimDist 2887 350
 

* While this column is capable of 200°C, it is designed for isothermal analysis at 110°C.