Maximum temperatures of packed columns – Porous Polymers
14 Jun 2015We often get asked about the maximum temperature limit of our packed columns, so I decided to write a post series which will provide this information for the majority of the packed columns which we sell. I thought I would start with the least complicated packed columns, those whose packings are typically not coated with a liquid phase. These packings include porous polymers.
For this post, I will only be focusing on HayeSep, Porapak, and Tenax packings. I decided not to include Chromosorb porous polymers because they are either in short supply, or currently unavailable.
I am going to make the following assumption: The column’s maximum temperature is limited to the packing, and not the column's tubing, end plugs, fittings/adaptors and/or ferrules used for installation. If you have a packed column which includes non-metal tubing or end-plugs other than glass wool or metal, contact Restek.
Please note: The catalog (part) numbers shown below are just for the packing, and not any packed column.
Keep watching for future posts on maximum temperatures of packed columns. Thank you.
HayeSep® Packing Materials
Description | temp. limits | g/btl. | Mesh 60/80 cat.# |
Mesh 80/100 cat.# |
Mesh 100/120 cat.# |
HayeSep A | 165 °C | 24 g | 22560 | 25032 | 25033 |
HayeSep B | 190 °C | 24 g | N/A | N/A | N/A |
HayeSep C | 250 °C | 24 g | N/A | 25036 | 25037 |
HayeSep D | 290 °C | 24 g | 25563 | 25038 | 25039 |
HayeSep DIP | 290 °C | 24 g | 25564 | N/A | N/A |
HayeSep DB | 290 °C | 24 g | N/A | 25568 | 25549 |
HayeSep DOX | (Use HayeSep DB) | ||||
HayeSep N | 165 °C | 24 g | 25570 | 25045 | 25046 |
HayeSep P | 250 °C | 24 g | 25571 | 25047 | 25048 |
HayeSep Q | 275 °C | 24 g | 25572 | 25049 | 25050 |
HayeSep R | 250 °C | 24 g | Not Available – Contact your local Restek Representative for suggested replacement. | ||
HayeSep S | 250 °C | 24 g | 25574 | 25053 | 25054 |
HayeSep T | 165 °C | 24 g | 25575 | 25005 | 25056 |
Porapak Packing Materials
Description | temp. limits | g/btl. | Mesh 50/80 cat.# |
Mesh 80/100 cat.# |
Mesh 100/120 cat.# |
Porapak P | 250 °C | 20 g | Available as Custom – Contact your local Restek Representative for ordering information. | ||
Porapak PS | 250 °C | 20 g | |||
Porapak Q | 250 °C | 26 g | |||
Porapak QS | 250 °C | 26 g | |||
Porapak R | 250 °C | 24 g | |||
Porapak S | 250 °C | 26 g | |||
Porapak N | 190 °C | 29 g | |||
Porapak T | 190 °C | 31 g |
Tenax® Packing Materials
Description | temp. limits | g/btl. | Mesh 60/80 cat.# |
Mesh 80/100 cat.# |
Tenax-TA | 350 °C | 10 g | 25550 | 25551 |
Tenax-GR | 350 °C | 10 g | 25552 | 25553 |