VICI Mat/Sen Purifier Module, Hydrogen, 1/8" Fittings
VICI Mat/Sen Purifier Module, Hydrogen, 1/8" Fittings
概要
- Replace separate traps for removing impurities (e.g., moisture and hydrocarbons) with one multiple-bed purifier that is designed specifically for purifying helium, hydrogen, nitrogen, or air.
- Reduce gas impurities from ppm to low ppb levels.
- Decrease baseline noise and increase GC-MS sensitivity.
- Pre-purged with the specified gas to shorten downtime.
Performance for these purifiers is optimized by incorporating a multiple-bed format that progressively lowers concentrations of contaminants at each successive bed. VICI Mat/Sen purifiers are guaranteed to produce gases that are purer than 99.9999%, when supplied with gas of 99.995% purity, and are pre-purged with the specified gas to speed conditioning. Purifier capacity is approximately four tanks of gas at 99.995% (50 ppm) purity, and correspondingly longer for purer gases.
Note: We recommend using an indicating oxygen trap (e.g., cat.# 22010) downstream from a VICI Mat/Sen purifier to continually ensure gas purity and indicate absolute changeout time.
Specifications
Length: 21" (53.3 cm)
Diameter: 1.5" (3.8 cm)
Maximum Inlet Pressure: 1000 psi (6895 kPa)
Maximum Recommended Flow: 500 mL/min
Pressure Drop from 120 psi (827 kPa) inlet at a flow of 0-500 mL/min: <0.20 psi (1.4 kPa)
Compression End Fittings: stainless steel
Shipping Weight: 3.04 lb (1300 g)