Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, Wall Mount or Bench Top
  1. Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, Wall Mount or Bench Top

Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, Wall Mount or Bench Top

Catalog No. 20684
Ordering Note
International power cords are available. Contact Customer Service to order.
Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, Wall Mount or Bench Top

Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, Wall Mount or Bench Top

Overview

Features & Benefits
  • Turn in-house compressed air into ultra-pure air.
  • Remove hydrocarbons to less than 0.1 ppm by catalytic oxidation.
  • Operate at 40–125 psi (276–862 kPa).
  • Typical payback is less than one year, based on cylinder costs.
  • Installs easily and takes up little bench space.
  • Maintenance kits include a one-year supply of prefilters and final filter.
Generator installation, warranty, and service plans are also available through Restek. Contact Restek Technical Service to arrange.
Description

Specifications for Parker Balston Zero Air Generators

Maximum Zero Air Flow Rate:75-83NA
HPZA-3500
HPZA-7000
HPZA-18000
HPZA-30000
1.0 L/min
3.5 L/min
7.0 L/min
18 L/min
30 L/min
Outlet Hydrocarbon Concentration (as methane):75-83NA
HPZA-30000
Other Models
< 0.1 ppm
< 0.1 ppm
< 0.05 ppm
Minimum/Maximum Inlet Air Pressure:40 psig/125 psig (276/862 kPa) 
Maximum Inlet Hydrocarbon Concentration (as methane):100 ppm 
Pressure Drop at Maximum Flow Rate:4 psi (28 kPa) differential 
Maximum Inlet Air Temperature:78°F (25°C) 
Inlet/Outlet Ports:1/4" NPT (female) 
Start-up Time to Specified Hydrocarbon Concentration:45 minutes 
Electrical Requirements:75-83NA
HPZA-3500, HPZA-7000
HPZA-18000, HPZA-30000
120 VAC/60 Hz, 0.5 amps
120, 220, 230, 240 VAC/60 Hz, 2.0 amps
120, 220, 230, 240 VAC/60 Hz, 4.0 amps
Dimensions:75-83NA
Other Models
10" h x 12" w x 3" d (25 cm x 30 cm x 8 cm)
16" h x 11" w x 13" d (42 cm x 27 cm x 34 cm)
Shipping Weight:75-83NA
Other Models
7 lb (3 kg)
41 lb (19 kg)

 

ModelNumber of FIDs (based on a 450 cc/min fuel air rate)
75-83NAUp to 2
HPZA-3500Up to 8
HPZA-7000Up to 16
HPZA-18000Up to 40
HPZA-30000Up to 66

Specifications

Product Specifications
Catalog No.
20684
Product Name
Parker Zero Air Generator, 75-83NA Model, 1000 cc/min Capacity, Wall Mount or Bench Top